Abstract: Objectives: To determine if point-of-care (POC) glycated haemoglobin (HbA1c) is sufficiently accurate in real-world remote settings to predict or exclude the diagnosis of diabetes based on laboratory HbA1c measurements. Design: Cross-sectional study comparing POC capillary HbA1c results with corresponding venous HbA1c levels measured in a reference laboratory. Participants: Aboriginal patients ≥15 years old who were due for diabetes screening at the participating clinics were invited to participate. Two hundred and fifty-five Aboriginal participants were enrolled and 241 were included in the analysis. Setting: 6 primary healthcare sites in the remote Kimberley region of Western Australia from September 2011 to November 2013. Main outcome measures: Concordance and mean differences between POC capillary blood HbA1c measurement and laboratory measurement of venous blood HbA1c level; POC capillary blood HbA1c equivalence value for screening for diabetes or a high risk of developing diabetes; sensitivity, specificity and positive-predictive value for diagnosing and screening for diabetes; barriers to conducting POC testing. Results: Concordance between POC and laboratory results was good (ρ=0.88, p<0.001). The mean difference was −0.15% (95% limits of agreement, −0.67% to 0.36%). POC HbA1c measurements ≥6.5%, 48 mmol/mol had a specificity of 98.2% and sensitivity of 73.7% for laboratory measurements ≥6.5%. The POC equivalence value for screening for diabetes or a high risk of developing diabetes was ≥5.7%, 39 mmol/mol (sensitivity, 91%; specificity, 76.7% for laboratory measurements ≥6.0%, 42 mmol/mol). Staff trained by other clinic staff ‘on the job’ performed as well as people with formal accredited training. Staff reported difficulty in maintaining formal accreditation. Conclusions: POC HbA1c testing is sufficiently accurate to be a useful component in screening for, and diagnosing, diabetes in remote communities. Limited local training is adequate to produce results comparable to laboratory results and accreditation processes need to reflect this.
Suggested Citation
Marley, JV, Oh, MS, Hadgraft, N, Singleton, S, Isaacs, K, Atkinson, D,
2015,
Cross-sectional comparison of point-of-care with laboratory HbA1c in detecting diabetes in real-world remote Aboriginal settings,
Volume:5, Journal Article,
viewed 11 October 2024,
https://www.nintione.com.au/?p=10976.